Trang chủ » IIoT » Bộ ghi dữ liệu CO2/Nhiệt độ/Độ ẩm/Dew Point từ xa với Ethernet/RS-485 và PoE ICP DAS DL-302-IP65 CR

Bộ ghi dữ liệu CO2/Nhiệt độ/Độ ẩm/Dew Point từ xa với Ethernet/RS-485 và PoE ICP DAS DL-302-IP65 CR

IP65 Remote CO2/Temperature/Humidity/Dew Point Data Logger with Safety Alarm (RS-485, Ethernet, PoE)

Chia sẻ:
Model: DL-302-IP65 CR
Views: 896

Bộ ghi dữ liệu CO2/Nhiệt độ/Độ ẩm/Dew Point từ xa với Ethernet/RS-485 và PoE ICP DAS DL-302-IP65 CR

The DL-300-IP65 series of Data Logger devices can be used to record CO, CO2, HCHO, TVOC, NH3, H2S, Temperature, Humidity and Dew Point information, including date and time stamps, and are able to store up to 450,000 downloadable records. Real-time data can be accessed from the DL-300-IP65 series Data Logger from anywhere and at any time using the free Windows software, the iOS App or the Android App, as long as they are connected to the same local network as the Data Logger. Support is provided for popular industrial protocols such as DCON, Modbus RTU, and Modbus TCP, as well as the emerging machine-to-machine (M2M)/IoT (Internet of Things) connectivity protocol-MQTT. The DL-300-IP65 series Data Logger can be connected via widely used communication interfaces including RS-485, Ethernet and PoE, meaning that the device can be easily integrated into existing HMI or SCADA systems, and is easy to be maintained in a distributed control system. The IP65 version of DL-300-IP65 series is designed for industrial applications in harsh environments that provides IP65 grade protection approval. The rugged RJ-45 ensures tight, robust connections, and guarantees reliable operation, even for applications that are subject to high vibration and shock.

Large 2.8″ LCD Touch Screen, with clear Color Chart to indicate the CO/CO2 Level

Supports the MQTT Protocol for IoT Applications

Simple and powerful utility
The DL300 Utility can be used to configure the modules, monitor real-time data, group DL-300 modules so that the status of distribution groups can be viewed and managed. The utility also allows the log data to be downloaded and exported to a .CSV file that can then be imported into any industrystandard software or spread sheet for analysis.

Multi-platform Remote Access Software
Real-time data from the DL-300 Data Logger can be accessed from anywhere and at any time using the DL300 Utility, the iOS or Android App, or via a regular web browser, as long as they are connected to the same local network as the Data Logger.

Display Messages in Multiple Languages

The display-message-on-screen function supports multiple language character sets based on UTF-8 encoding. Either pre-configured messages or dynamic messages can be remotely displayed using Modbus commands, or a dynamic message can be sent via the web-based interface.

Mushroom Farm
Mushroom growth yield requires long time to monitor the environment of CO2 concentration, temperature and humidity values, and its growth environment is high humidity condition. The general equipment cannot work in such environment, but the DL-302-IP65 can be used to record CO2, temperature, humidity and dew point information with date and time stamps in the operating temperature from -20˚C to +50˚C, and are able to store up to 450,000 downloadable records. The WISE-5231-3GWA features timer and schedule functions. It allows user to schedule specific date or time for control logic execution, or perform specific tasks such as time delay. With calendar user interface provided, schedule setting can be more efficient and flexible. This does not require programming, only some simple settings will be able to achieve the farm production control and quality control, production resume … and so on.Pig Farm
Ammonia and Hydrogen sulfide are two of the most harmful gases in animal farming environment. They will directly affect animals’ growth and immunity. Pig farm requires long-term immediate monitoring the environment of NH3 concentration, temperature and humidity values. The general equipment cannot work in such environment, but the DL-305-IP65 can be used to record NH3, temperature, humidity and dew point information with date and time stamps in the operating temperature from -20°C to +50°C, and are able to store up to 450,000 downloadable records. The WISE-5231-3GWA features timer and schedule functions. It allows user to schedule specific date or time for control logic execution, or perform specific tasks such as time delay. With calendar user interface provided, schedule setting can be more efficient and flexible. This does not require programming, only some simple settings will be able to achieve the farm production control and quality control, production resume … and so on.

Models DL-301-IP65 DL-302-IP65 DL-303-IP65 DL-304-IP65 DL-305-IP65 DL-306-IP65 DL-307-IP65 DL-308-IP65
System
CO Alarm Yes Yes
CO2 Alarm Yes Yes
HCHO Alarm Yes Yes
TVOC Alarm Yes Yes
NH3 Alarm Yes
H2S Alarm Yes
RH/Temperature Alarm Yes
Real Time Clock Yes
Data Logger Yes, up to 450,000 records
Relay Output PhotoMos Relay × 4, 100 VDC @ 1 A
Communication
RS-485 Port Baud Rate = 1200 ~ 115200 bps
Ethernet Port 10/100 Base-TX, 8-Pin RJ-45 x1(Auto-negotiating, Auto-MDI/MDIX, LED indicators)
Security IP filter (whitelist) and Password (web)
Protocol Modbus/RTU(RS-485), Modbus TCP(Ethernet) and MQTT(Ethernet)
Dual Watchdog Yes, Module (2.3 sec), Communication (Programmable)
LED Indicators
PWR Green for normal operation
Link Green for the Ethernet-linked
S1 ~ S4 Red for an alarm condition
Main Machine Interface
LCD 2.8 TFT (Resolution 240 x 320 x 16), Defective Pixels <= 3
Backlight Life 20,000 hours
Brightness 160 cd/m2
Touch Panel Yes
Electrical
Powered from Terminal Block +12 to +48 VDC
Powered from PoE IEEE 802.3af, Class 1 (48 V)
Power Consumption PoE 2.9 W (Max.) 3.0 W (Max.) 3.1 W (Max.) 2.0 W (Max.) 2.9 W (Max.) 2.9 W (Max.) 2.1 W (Max.) 2.0 W (Max.)
Non-PoE 2.7 W (Max.) 2.8 W (Max.) 2.8 W (Max.) 2.0 W (Max.) 3.0 W (Max.) 3.0 W (Max.) 2.1 W (Max.) 2.0 W (Max.)
Mechanical
Installation DIN-Rail or Wall Mounting
Dimensions (L x W x H) 170 mm x 117 mm x 60 mm
Environment
Operating Temperature -20 to +50°C
Storage Temperature -30 to +75°C
Humidity 10 to 90% RH, Non-condensing
Protection Class IP65
Models DL-301-IP65 DL-302-IP65 DL-303-IP65 DL-304-IP65 DL-305-IP65 DL-306-IP65 DL-307-IP65 DL-308-IP65
CO Measurement
Range 0 to 1000 ppm
(Electrochemical)
0 to 1000 ppm
(Electrochemical)
Resolution 1 ppm 1 ppm
Accuracy ±5% of
measured value
±5% of
measured value
Response Time 30 sec 30 sec
Warm-up Time 300 sec 300 sec
CO2 Measurement
Range 0 to 9999 ppm (NDIR)
Resolution 1 ppm
Accuracy ±30 ppm ±3% of measured value
Response Time 20 sec
Warm-up Time 60 sec
HCHO Measurement
Range 0 ppb to 2000 ppb
(Electrochemical)
0 ppb to 2000 ppb (Electrochemical)
Resolution 1 ppb 1 ppb
Accuracy 0 ~ 300ppb :
±30ppb >
300ppb : ±10%
0 ~ 300ppb :
±30ppb >
300ppb : ±10%
Response Time ≤60 sec ≤60 sec
Warm-up Time 180 sec 180 sec
TVOC Measurement
Range 0 ppb to 60000
ppb (MEMS
Metal Oxide)
0 ppb to 60000
ppb (MEMS
Metal Oxide)
Resolution 1 ppb 1 ppb
Accuracy ±15% ±15%
Response Time 60 sec 60 sec
Warm-up Time 180 sec 180 sec
NH3 Measurement
Range 0 to 100 ppm
(Electrochemical)
Resolution 1 ppm
Accuracy ±5% of
measured value
Response Time < 40 Sec
Warm-up Time 60 Sec
H2S Measurement
Range 0 to 100 ppm
(Electrochemical)
Resolution 1 ppm
Accuracy ±5% of
measured value
Response Time < 30 Sec
Warm-up Time 60 Sec
Temperature Measurement
Range -20 to +50°C
Resolution 0.1°C
Accuracy ±0.6°C
Relative Humidity Measurement
Range 0 to 100% RH, Non-condensing
Resolution 0.1% RH, Non-condensing
Accuracy ±5% RH, Non-condensing
Dew Point
Range Calculated using temperature and relative humidity
Resolution 0.1°C
DL-302-IP65     CR IP65 Remote CO2/Temperature/Humidity/Dew Point Data Logger with Safety Alarm (RS-485, Ethernet, PoE) (RoHS)