Trang chủ » IIoT » IO/Sensor » Industrial Sensor » Cảm biến đo và lưu trữ O2/CO/Nhiệt độ/Độ ẩm/Điểm sương từ xa ICP DAS CL-251-E CR

Cảm biến đo và lưu trữ O2/CO/Nhiệt độ/Độ ẩm/Điểm sương từ xa ICP DAS CL-251-E CR

Remote O2/CO/Temperature/Humidity/Dew Point Data Logger Module

Chia sẻ:
Model: CL-251-E CR
Views: 1172

Cảm biến đo và lưu trữ O2/CO/Nhiệt độ/Độ ẩm/Điểm sương từ xa ICP DAS CL-251-E CR

The CL-25X-E series of O2 and gas measurement module that measures oxygen gas (O2) in air. In addition, various fume concentrations related to human health can also be measured. For example: CO / CO2 / temperature / humidity and dew point information, including the date and time stamps, and are able to store up to 450,000 downloadable records. Real-time data can be accessed from the CL-200 data logger from anywhere and at any time using the free Windows software, the Iso App, or the Android App, as long as they are connected to the same local network as the data logger. Support is provided for common industrial protocols such as DCON, Modbus RTU, and Modbus TCP, as well as the emerging machine to-machine (M2M)/)IoT (Internet of Things) connectivity protocol – MQTT. The CL-200 data logger can be connected via widely used communication interfaces including RS-485, Ethernet and PoE, meaning that the device can be easily integrated into existing HMI or SCADA systems, ensuring trouble-free maintenance in distributed control systems

Supports the MQTT Protocol for IoT Applications

Multi-platform Remote Access Software

Simple and Powerful DL-300 Utility

 

Temperature Measurement
Range -10 to +50°C
Resolution 0.1°C
Accuracy ±0.6°C
Humidity Measurement
Range 0 to 100% RH, Non-condensi
Resolution 0.1% RH, Non-condensing
Accuracy ±5% RH, Non-condensing
Life Time 10 years
Dew Point
Range Calculated using temperature and relative humidity
Resolution 0.1°C
O2 Measurement
Range 0~25%
Resolution 0.01%
Accuracy 2% of FSR
Response Time < 30 secs (typical)
Warm-up Time 120 seconds
Life Time > 5 years
Re-calibratable No
Offset Programmable Yes
Replaceable (RMA) Yes
CO Measurement
Range 0 to 1000 ppm (Electrochemical)
Resolution 1 ppm
Accuracy ±5% of measured value
Response Time 30 seconds
Warm-up Time 60 seconds
Life Time 5 years
Re-calibratable No
Offset Programmable Yes
Replaceable (RMA) Yes
System
Alarm O2/CO/RH/T
Real Time Clock Yes
Data Logger Yes, up to 450,000 records
Relay Output
Channels 1 x Form A, SPST
30 VDC @ 16 A or 250 VAC @ 16 A
COM Ports
Ports 1 x RS-485
1 x Ethernet
Protocol DCON and Modbus/RTU (RS-485)
Modbus TCP and MQTT (Ethernet)
Ethernet
Ports 10/100 Base-TX, 8-Pin RJ-45 x1
(Auto-negotiating, Auto-MDI/MDIX, LED indicators)
Power
Consumption Non-PoE: 2.2 W (Max.)
PoE: 2.4 W (Max.)
Powered from Terminal Block +10 ~ +30 VDC
Mechanical
Dimensions (mm) Ø 150 x 53 (Ø x H)
Installation Ceiling Mounting/Wall Mounting
Ingress Protection Rating IP20
Environmental
Operating Temperature 0 ~ 50°C
Storage Temperature -30 ~ +75°C
Humidity 10 ~ 90% RH, Non-condensing
CL-251-E     CR Remote O2/CO/Temperature/Humidity/Dew Point Data Logger Module (RoHS)